Maximum contact surface temperature 44.1° C (111.4° F) Average time for contact surface temperature to rise from 23° ± 2° C to 37° C is 6 minutes Average time for temperature of air exiting the hose to rise from 23° ± 2° to 37° C: <1 minute Accuracy of displayed...
- Maximum contact surface temperature 44.1° C (111.4° F)
- Average time for contact surface temperature to rise from 23° ± 2° C to 37° C is 6 minutes
- Average time for temperature of air exiting the hose to rise from 23° ± 2° to 37° C: <1 minute
- Accuracy of displayed temperature: ±1.0°C (air entering hose)
- After 45 minutes of continuous use, blower will step down from the boost to high setting
- Thermostat (internal): 49°C to 55°C (120°F to 131°F)
- Average alarm level: 56 dB @ 1 meter
- 100 to 240 V AC, 8 A max. current at 100 V, 5 A max. current at 240 V power requirements
- Input frequency: 50/60 Hz ±1 Hz
- Fuse (x2): little fuse (mfr.) 0218010; 250 V, 10 A
- 18°C to 28°C (64.4°F to 82.4°F) Temperature